85
Responding to Accuracy Range Alarm
The Camino ICP Monitor is designed to measure ICP and tissue temperature between specific clinical ranges:
Measurements
Range
ICP
Between -10 to +125 mmHg that includes the following levels of accuracy
depending on catheter type:
Fiber Optic Catheter:
•
-10 to 50 mmHg (±2 mmHg)
•
51 to 125 mmHg (±6%)
Flex Catheter:
•
-10 to 10 mmHg (±2 mmHg)
•
11 to 33 mmHg (±3 mmHg)
•
34 to 125 mmHg: ± (6% + 1 mmHg)
Temperature
Between 30 °C to 42 °C (±0.3 °C)
If the measured ICP or temperature falls outside these ranges, the monitor will activate a low priority alarm that
will:
•
Sound an alarm containing a burst of one single pulse every 16 seconds
•
Display one of the following messages on the status bar:
"
ICP is out of accuracy range"
or "
Temperature is out of accuracy range"
•
Display the Mean ICP or temperature value in yellow on the
Main
,
Trend
, and
Alarm
screens
To resolve this problem, the Mean ICP or temperature must fall within the rated accuracy ranges listed in the
above table.
Silencing the ICP and Temperature Accuracy Range Alarm Temporarily
Press the yellow alarm symbol on the touch screen. This will silence the alarm for 3 minutes while the Mean ICP or
temperature values fall back within the rated accuracy ranges displayed in the above table. If the Mean ICP or
temperature values remain outside the rated accuracy ranges for the 3 minutes, the alarm will reactivate.
.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...