140
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
Power
(W)
Distance
(m)
Immunity Test level
(V/m)
385
380-390
TETRA 400
Pulse Modulation
b
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
c
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
745
780
704 – 787
LTE Band 13,
17
Pulse modulation
b
217 Hz
0.2
0.3
9
810
870
930
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
b
18 Hz
2
0.3
28
1720
1845
1970
1700 – 1990 GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modulation
b
217 Hz
2
0.3
28
2450
2400 – 2570 Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
b
217 Hz
2
0.3
28
5240
5500
5785
5100 – 5800
WLAN 802.11
a/n
Pulse modulation
b
217 Hz
0.2
0.3
9
NOTE
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...