51
Procedures for Synchronizing the Two Monitors
Perform the following steps to verify that both the Camino ICP Monitor and the patient bedside monitor report
the same ICP values, ±1 mmHg or 1 % of monitor reading (whichever is greater).
Note that these steps should be performed in each of the following situations:
•
When first connecting the Camino ICP Monitor to a patient bedside monitor.
•
If the Camino ICP Monitor becomes disconnected from a patient bedside monitor (e.g. due to patient transport
to CT or OR), and then needs to be re-connected.
•
If, during use, the reported ICP values between the Camino ICP Monitor and the patient bedside monitor differ
by more than ± 1 mmHg or 1 % of monitor reading (whichever is greater).
If the ICP value is above 100 mmHg, the difference between both monitors can differ by 2 mmHg.
Step 1: Connect the Two Monitors
Notice
To cancel the synchronization process during any of the following steps, press
Done
.
Connect the Camino ICP Monitor to the patient bedside monitor: