90
Extract Data via Digital Streaming
Use the streaming data feature to export Mean ICP trend data digitally from the Camino ICP Monitor to a remote
device via a USB-to-RS232 adapter cable. Note that the Natus Neuro supplied USB-to-RS232 adapter cable is
required to use this feature. When streaming data, the monitor reports the mean ICP value in mmHg of the current
ICP trend data in one-minute intervals. The USB-to-RS232 adapter cable delivers the information as 9600 baud,
8 data bits, 1 stop bit, even parity configurations.
To connect the USB-to-RS232 adapter cable to a computer, you need a female to female, null modem RS232
cable. To extract the data via digital streaming:
1.
On the rear panel, attach the USB end of the USB-to-RS232 adapter cable to the
USB
connector.
1
USB-to-RS232 adapter cable
2
USB end of adapter cable
2.
On the touch screen, press
Trend
and verify the following message appears on screen: "
data
streaming enabled"
3.
Connect the other end of the USB-to-RS232 adapter cable to the external monitoring device.