103
System Message
Cause
Recommendation
More than one catheter
connected.
Two ICP catheters types (Fiber Optic and Flex)
are connected simultaneously.
Remove the ICP catheter that will not be used during patient
treatment.
No catheter is connected
No ICP catheter is connected to Camino ICP
Monitor.
Connect ICP catheter (see Chapter 3).
Temperature is out of
accuracy range
Temperature value has fallen under range that
the Camino ICP Monitor can measure accurately
(between 30 °C to 42 °C).
Do not rely on temperature measurements until the
temperature value falls within the accuracy range.
Temperature sensor
failure
Temperature catheter broken.
Disconnect temperature catheter and replace with new one
(see directions for use supplied with catheter).