Sterilizing the Camino
Fiber Optic Catheter Cable
The Fiber Optic Cable (REF # CAMCABL) is provided non-sterile.
Notice
The Fiber Optic Catheter Cable must be sterilized whenever it is to be used in the sterile field. Prior to sterilization, clean
this cable using the guidelines listed in Chapter 7.
Limits on Sterilization
The Fiber Optic Catheter Cable may be sterilized thirty (30) times and cleaned/ disinfected one hundred (100)
times. Do not use the cable after reaching these reprocessing limits. For tracking purposes, use the
reprocessing record sheet (doc # 025471) supplied with your cable
Sterilization Parameters
The Camino Fiber Optic Catheter Cable supports EtO and Sterrad sterilization methods.
Notice
Prior to EtO sterilization, the cable should be individually packaged in a self-sealing sterilization pouch with dimensions 10" x
15" (or equivalent pouches designed for EtO gas sterilization).
Notice
Prior to Sterrad sterilization:
The cable should be placed in a 12" x 15" Tyvek® Self-Seal Pouch (or equivalent pouches designed for Sterrad sterilization);
- or -
The cable should be placed in a 23" x 11" x 4" Sterrad® compatible Aptimax® tray (or an equivalent instrument tray
designed for Sterrad sterilization) and wrapped in Kimguard® One-Step® Kc400 Sterilization Wrap, then secured with
Sterrad compatible Sealsure® chemical indicator tape (or equivalent sterilization wrap and indicator tape designed for
Sterrad sterilization)
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...