80
Responding to the Physiological Alarm (ICP above alarm limit)
If the patient’s Mean ICP value exceeds the specified limit for more than 5 seconds, the Camino ICP Monitor will
activate a medium priority physiological alarm that will:
•
Sound an alarm containing a burst of three pulses
•
Display the following error message on the status bar: "
ICP above alarm limit"
•
Flash the alarm symbol on status bar
•
Flash the Mean ICP value in yellow on the
Main
,
Trend
, and
Alarm
screens
To silence the alarm temporarily, press the flashing alarm symbol on the touch screen. This will silence the alarm
for either 3 minutes or until the patient’s Mean ICP value falls within the specified alarm limit for at least one
measurement. After 3 minutes, the alarm will reactivate. To turn off the alarm completely, press the
Alarm
tab and
select
Alarm Off
and
Accept
. For instructions on specifying the high ICP alarm limit.
Warning
Selecting the
Alarm Off
feature will disable the high ICP alarm limit indefinitely. Use caution if this feature is selected.