125
7.
On the touch screen, press the
Main
tab and verify the
ICP
value displays approximately 10 ± 2 mmHg.
After about 5 seconds, verify the monitor:
•
Sounds an alarm containing a burst of three pulses
•
Displays the following error message on status bar: "
ICP above alarm limit"
•
Flashes the alarm symbol on status bar
•
Flashes the current ICP value in yellow on the
Main
,
Trend
, and
Alarm
screens
8.
Silence the alarm temporarily. On the touch screen, press the flashing alarm symbol and verify the
monitor:
•
Silences the alarm.
•
Changes the active alarm symbol to the silenced alarm symbol.
•
Displays
Audio Paused
on the status bar.
9.
Wait approximately 3 minutes and verify the monitor re-activates the alarms specified in step 7.
10.
Decrease simulated pressure to 0 mmHg, and verify the alarms turn off.
11.
On the touch screen, press the
Alarm
tab and select
Alarm Off
.