Guidance and manufacturer’s declaration - electromagnetic emissions
The Natus Camino ICP Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Natus
Camino ICP Monitor should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Natus Camino ICP Monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The EMISSIONS characteristics of Camino ICP Monitor make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) Camino ICP
Monitor might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies