114
5.
Insert the tip of the Fiber Optic Catheter into the Y-hemostasis valve through the duckbill valve.
6.
Open the valve from the I.V. bag and remove the end of the cap from the Y-hemostasis valve. Let water flow
until the tubing and Y-hemostasis valve fill with water.
7.
Verify that air bubbles are not present and replace the end cap.
8.
Verify that the water level from the I.V. bag and the tip of the Fiber Optic Catheter are at the same level. In
this example, the catheter’s tip and the water are leveled with the 10-inch mark of the ruler.
1
Water level and catheter tip
2
End Cap
9.
Turn on the monitor. On the touch screen, press the
Main
tab and verify the ICP value reads 0. If not, use the
black zero adjustment tool from the catheter kit to set the ICP value to 0 mmHg.