89
4.
Press
Accept
, remove the USB drive, and view the
.csv
file in the desired output device (e.g. laptop).
Notice
The USB drive needs to be a FAT (FAT16 or FAT32) formatted drive. Other file system formats are not supported.
Notice
To ensure the complete transfer of trend data to the USB drive, do not remove the USB drive until the message specified in step 3
appears confirming the extraction.
Notice
In the
.csv
file, the timestamp for the data extraction is based on the monitor’s date/time settings
that are specified on the
Settings
tab. In countries where daylight savings are observed, the time
must be adjusted manually. For instructions, see Chapter 2.
How the Monitor Stores Trend Data for Up to 5 Days
Warning
The Camino ICP Monitor will only store the Mean ICP data from the most recent 5 days. All stored trend data older than 5 days will
be lost. If monitoring is continued for more than 5 days, placement of a new catheter under sterile conditions is recommended. Note
that replacing a catheter will reset the trend data. Please extract any data that you wish to retain prior to replacing the catheter.