66
Specifying the High ICP Alarm Limit
Warning
Always verify the high ICP alarm limit is set appropriately for each patient prior to treatment.
1.
On the touch screen, press the
Alarm
tab.
2.
Press the
Alarm On
button; this will activate the alarm settings.
3.
In the
High ICP Alarm Limit
field, use the arrows to specify the high ICP alarm limit value for the
patient. Note that the range limits are between -10 to +125 mmHg in 1 mmHg increments.
4.
Press
Accept
. If the patient’s Mean ICP value exceeds this specified limit for more than 5 seconds, the
monitor will activate the physiological alarm. Note that pressing
Cancel
will revert all changes back to
the previous values. To respond to this alarm, see Chapter 4.