Page 90
Appendix |Multilayer Measurement, Controller IFC24x1MP
confocalDT 24x1
A 4.2
LED Meaning for Multilayer Measurements
Name
State
Description
LED 1
Intensity
Red
Intensity too high, at least one of the peaks is in satura-
tion.
Yellow
Intensity too low, at least one of the peaks is below the
recognition threshold.
Green
Signal ok
LED 2
Range
Red
Focus of too less (<N) peaks within the linearized (or
masked) range.
Yellow
Average value from first and nth peak in the centre of
measuring range (47.5 … 52.5 %)
Green
OK, all used peaks are within the linearized (or masked)
range.
Error 1
Intensity = 1
Warning if the intensity of one or more peaks falls within
saturation or below the recognition threshold.
Error 2
Out of range = 1 Too less (<N) peaks within the linearized area.
N - Number of (expected) peaks
A 4.3
Digital Interfaces / Data Selection
Here the data which should be transferred via the digital interfaces can be selected.