Page 52
Advanced Settings
confocalDT 24x1
6.
Advanced Settings
6.1
Login, Switching User Level
Assigning passwords prevents unauthorized changes to controller settings. Password
protection is not enabled as a factory setting. The sensor works on the
User
level. After
the controller has been configured, you should enable password protection. The stan-
dard password for the User level is “000”.
i
A software update will not change the standard password or a custom password.
The user level password is setup-independent, and is not loaded or stored during
setup.
User can do the following:
User
Professional
Password required
no
yes
View settings
yes
yes
Change settings, change passwords
no
yes
View readings, video signals
yes
yes
Scale graphs
yes
yes
Restore factory settings
no
yes
Fig. 37 Permissions within the
user
hierarchy
Enter the standard password
“000” or a custom password into
the
Password
box, and click
Login
to confirm.
To change to user mode, click the
Logoff
button.
Fig. 38 Changing to professional level
In
User
mode, you can use the change password features to assign a custom pass-
word.
Password
Value
All passwords are case-sensitive. Numbers are al-
lowed, but special characters are not permitted.
User level at
restart
User /
Professional
Defines the
user
level that is enabled when the sensor
starts the next time. MICRO-EPSILON recommend to
select Professional level.
i
When an professional restores the factory settings (Settings menu > Tools > Fac-
tory Settings), the Professional level password is reset to “000”.
6.2
Detection Threshold
The detection threshold (in %, relates to the signal after dark correction) defines the mini-
mum intensity for including a video signal peak in the measurement. Therefore, the video
graph must be taken into consideration when defining the threshold.
Peak detection threshold
Value
Value in %, factory setting: 1%
Defining the peak detection threshold
-
For very weak signals (e.g. typical for extremely high measuring rates), choose a low
detection threshold, as only signal parts above this threshold will be included in mea-
surements.
-
For thickness measurements, you may increase the detection threshold, if video signal
peaks merge. In general, set the threshold high enough to avoid that any interfering
video signal peaks are detected.
The detection threshold affects linearity, it is therefore recommended to adjust it as little
as possible.
Click
Submit
to confirm selection of the recognition threshold.
Fields with a grey back-
ground require a selection.
Value
Fields with a dark border
require the specification of
a value.
MP
Parameter available with
controller IFC24x1MP