A 5.3.3.6 Light Correction
LIGHTCORR
Performs a light correction. The correction is light source-dependent not sensor-depen-
dent. The light correction is already performed by the manufacturer.
For the IFC with xenon light source can this correction be performed by the user after a
light source change in the user level PROFESSIONAL. The IFC2451, IFC2461 and the
IFC2471LED require no light correction by the customer.
Error
Description
E04
Error with the setting of internal parameters (should never occur)
E16
Timeout when light correction
E18
A data transmitting is already running -> Light correction not possible
A 5.3.4 Trigger Modes
Trigger-input serves also as synchronous input, which means level and edge triggering is
only alternatively possible to sync mode.
A 5.3.4.1 Trigger Type
TRIGGER NONE|EDGE|PULSE|SOFTWARE|ENCODER
-
NONE: No triggering
-
PULSE: Level triggering
-
EDGE: Edge triggering
-
SOFTWARE: Software triggering
-
ENCODER: Encoder triggering
A 5.3.4.2 Effect of the Trigger Input
TRIGGERAT [INPUT|OUTPUT]
INPUT: Triggering the measured value recording. When calculating the mean, mea-
sured values immediately before the trigger event are not included; instead
older measured values are used, which were the output in previous trigger
events.
OUTPUT: Triggering the measurement value output. When calculating the mean, mea-
sured values immediately before the trigger event are used.
Triggering of measured value recording is enabled as a factory default setting.
A 5.3.4.3 Trigger Level
TRIGGERLEVEL HIGH|LOW
-
HIGH: Edge triggering: Rising edge, level triggering: High-active
-
LOW: Edge triggering: Falling edge, level triggering: Low-active
A 5.3.4.4 Number of Output Measurement Values
TRIGGERCOUNT <1...16382>|16383
Number of measurement values which are displayed when edge triggering or software
triggering.
-
1...16382: Number of measurement values which are displayed after a trigger impulse
when edge triggering or software triggering.
-
16383: Start infinite output of measurement values after a trigger impulse when edge
triggering or software triggering.
-
0: Stop triggering