Page 131
Appendix |EtherCAT Documentation
confocalDT 24x1
Objekt 2156h: Multilayer measurement options
2156
RECORD Multilayer options
ro
Subinzes
0
VAR
Number of entries
2
Unsigned8
ro
1
VAR
Number of peaks for
multilayer
2
Unsigned8
rw
2
VAR
Disable refractivity cor-
rection
FALSE
BOOL
rw
Further details can be found in the chapter multilayer measurement, see Chap.
.
Refrac count:
Number of processed distances (2 ... 6)
Disable refractivity correction:
Deactivation of the refractive index correction
Objekt 2158h: Thickness measurements options
2158
RECORD Thickness options
ro
Subinzes
0
VAR
Number of entries
1
Unsigned8
ro
1
VAR
Disable refractivity cor-
rection
FALSE
BOOL
rw
Further details can be found in the chapter multilayer measurement, see Chap.
.
Disable refractivity correction:
Deactivation of the refractive index correction
Objekt 2161h: Peak position
2161
VAR
Peak position
0
Unsigned8
rw
The peak/peaks that is/are evaluated in the distance or thickness measurement mode
can be set using this command.
Standard: first peak / first and second peak
In order to get a reproducible measurement result, the default setting should be changed
only in urgent cases.
Position for displacement
Position for thickness measurement
0
first peak
0
first and last peak
1
last peak
1
second to last and last peak
2
first peak
2
first and second peak
3
highest peak
3
highest and second highest peak