Page 39
Operation
confocalDT 24x1
5.3
User Interface, Basic Preferences
5.3.1
Introduction
The following chapters describe any controller settings that are required to get started
and quickly achieve first measuring results.
You can access additional submenus, e.g. for measuring rates and triggers, through the
navigation bar on the left side of a web page.
i
When programming has been completed, all settings must be permanently stored
in a set of parameters to ensure that these settings will be available when the sen-
sor is switched on the next time.
Some parameters are only
available in the controller
IFC24x1MP. This is char-
acterized by a separate
formatting.
Distance measurement
Parameter available with
controller IFC24x1 and con-
troller IFC24x1MP
Thickness measurement
Multilayer measurement
Parameter available with
controller IFC24x1MP
For details about other measuring control settings and features, such as setting masters
or triggers, see “Advanced settings”, see Chap.
5.3.2
Measurement Program
Select the desired type of measurement program from the list:
confocal
DT
2451
Measurement program
Login
Measurement program
Sensor
Averaging/error handling
Zeroing/mastering
Exposure mode/measuring rate
Detection threshold
Material database
Switching outputs
Analog output
Digital interfaces
Output-data rate
Encoder inputs
Trigger mode
Synchronization
Distance Measurement
Submit
Home
Preferences
Measurement Video signal
Help/Info
Preferences > Measurement program
Measurement to be effected
First peak
Used peak
Distance measurement
Distance to the first surface
Thickness measurement
Thickness of one layer - distance between the first two peaks
Status:
OK
Save Setup
Type of
measure-
ment
Distance
Used peak
first / highest / last peak
Distance to n-th surface,
depends on the selected
peak
Thickness
Used peak
first and second /
first and last /
second to last and last /
highest and second high-
est peak
One-sided thickness
measurement of transpar-
ent materials; the relevant
material needs to be se-
lected (refractive index).
Selection
of material
Vacuum / water ...
Multilayer measurement
Selectable distances for
up to 6 peaks
Use refractive correction:
Yes / No
Material between Peak 1 /
2 ... 6 and Peak 2 / 3 ... 6
i
The selected measuring program is used as the standard measuring program on
startup.
Fields with a grey back-
ground require a selection.
Value
Fields with a dark border
require the specification of
a value.
MP
Parameter available with
controller IFC24x1MP