Page 40
Operation
confocalDT 24x1
5.3.3
Material Database
Select Material
One-sided thickness measurements require the refractive index of the transparent mate-
rial to calculate the actual thickness. For this purpose, different material data are stored
in the controller.
You can select the transparent material in the Thickness Measurement program, see
.
Select the transparent material to be used for measuring the thickness.
Click
Submit
to confirm your selection.
The controller stores a material table that can be modified and added to, see Chap.
confocal
DT
2451
Login
Measurement program
Sensor
Averaging/error handling
Zeroing/mastering
Exposure mode/measuring rate
Detection threshold
Material database
Switching outputs
Analog output
Digital interfaces
Output- data rate
Encoder inputs
Trigger mode
Synchronization
Home
Preferences
Measurement
Video signal
Help/Info
PS
Polystyrol,
a plastic
1.604079
1.590481
1.584949
PMMI
Polymethacryl
methylimid,
a plastic
1.534000
1.534000
1.534000
Material database
Preferences > Material database
Material name Description
Vacuum, Air
Vacuum; air
1.000000
1.000000
1.000000
Fused Silica
1.463126
1.458464
1.456367
Refractive index
nF at 486 nm
Refractive index
nD at 587 nm
Refractive index
nC at 656 nm
Abbe value nd
BK7
1.522380
1.516800
1.514320
PMMA
Acrylic glass
1.497761
1.491756
1.489200
Acrylic
Acrylic rosin, e.g
adhesive, lacquer
Crown glass
Quartz glass,
silica
1.497828
1.491668
1.488938
PC
Polycarbonat,
a plastic
1.599439
1.585470
1.579864
Save Setup
Micro-Epsilon assumes no responsibility for the stored material parameters.
5.3.4
Selecting a Sensor
Controller and sensor(s) are matched at the factory.
Select the connected sensor (type, range and serial number) from the list.
Click
Submit
to confirm your selection.
confocal
DT
2451
Sensor
Login
Measurement program
Sensor
Averaging/error handling
Zeroing/mastering
Exposure mode/measuring rate
Detection threshold
Material database
Switching outputs
Analog output
Digital interfaces
Output-data rate
Encoder inputs
Trigger mode
Synchronization
IFS2405-1 SN71011
Submit
Home
Preferences
Measurement
Video signal
Help/Info
Preferences > Sensor
Connected Sensor:
Status:
OK
The connected sensor is to be chosen from the stored calibration data (according to type, measurement
range and serial number).
Calibration data of up to 20 different sensors can be stored within the controller.
Save Setup
The calibration data of up to 20 different sensors can be stored in the controller.