Page 162
Appendix |Operating Menu
confocalDT 24x1
A 7
Operating Menu
Login
Announced user level
Value
only reading
Change to user- / professional
level
Button logoff or enter password
Change password
Old password
Value
New password
Value
Repeat new password
Value
User level by restart
Professional / User
Type of measure-
ment
Distance
measurement
Used peak
first / highest / last peak Distance measurement: Distance
to the n-th surface, depends on
the selected peak.
Thickness
measurement
Used peak
f
irst and second /
first and last /
second to last and last /
highest and second
highest peak
One-sided thickness measure-
ment of transparent materials;
the relevant material needs to be
selected (refractive index).
Selection of material
Vacuum / water ...
Multilayer Measurement
1)
Multilayer Measurement: Select-
able distances for up to 6 peaks
Use refractive correction: Yes / No
Material between Peak 1 / 2 ...6
and Peak 2 / 3 / ... 6
Sensor
Sensor type and serial number
Controller stores max. 20 different sensors.
Exposure mode/
measuring rate
Exposure mode
Automatic mode / Measurement mode / Manual mode /
Alternating Two-Time-Mode / Automatic Two-Time-Mode
Measuring rate
-
with IFC2451: 0.1 / 0.2 / 0.3 / 1 / 2.5 / 5 / 10 kHz
-
with IFC2461: 0.1 / 0.2 / 0.3 / 1 / 2,5 / 5 / 10 / 25 kHz
-
with IFC2471LED: 0.1 / 1 / 2.5 / 5 / 10 / 25 / 50 / 70 kHz
-
with IFC2471: 0.3 / 1 / 2.5 / 5 / 10 / 25 / 50 / 70 kHz
Exposure time 1 in µs
Value (0.1 µs ... 10.000 µs (IFC2451))
Value (0.1 µs ... 10.000 µs (IFC2461))
Value (0.1 µs ... 3333.3 µs (IFC2471))
Exposure time 2 (shorter) in µs
Value (Value smaller than Exposure time 1)
Detection
threshold
Value
Value in %, factory setting 1 %
1)
Only with controller IFC24x1MP available