Appendix |ASCII Communication with Controller
confocalDT 24x1
A 5.5.3.3 Switching Level
ERRORLEVEL HIGH|LOW
-
HIGH: Switching output is High upon error
-
LOW: Switching output is Low upon error
A 5.5.4 Analog Output
A 5.5.4.1 Data Selection
Controller IFC24x1
Controller IFC24x1MP
ANALOGOUT DIST1|DIST2|THICK12
ANALOGOUT DIST1|DIST2|DIST3|
DIST4|DIST5|DIST6|THICK12|
THICK13|THICK14|THICK15|
THICK16|THICK23|THICK24|
THICK25|THICK26|THICK34|
THICK35|HICK36|THICK45|
THICK46|THICK56
Selection of the signal which should be output via the analog output.
DIST1: Output of displacement 1
-
DIST2: Output of displacement 2
-
THICK12: Output of the difference between displacement 1 and 2
A 5.5.4.2 Output Range
ANALOGRANGE NONE|0-5V|0-10V|-5-5V|-10-10V|4-20mA
-
NONE: No analog output (inactive)
-
0 - 5 V: The analog output outputs a voltage between 0 up to 5 volt.
-
0 - 10 V: The analog output outputs a voltage between 0 up to 10 volt.
-
-5 - 5 V: The analog output outputs a voltage between -5 up to 5 volt.
-
-10 - 10 V: The analog output outputs a voltage between -10 up to 10 volt.
-
4 - 20 mA: The analog output outputs an intensity of current of 4 up to 20 milliam-
peres.
A 5.5.4.3 Two-Point Scaling
ANALOGSCALE STANDARD|(TWOPOINT <Minimum measured value> <Maximum
measured value>)
Setting the scaling of analog output.
The default scaling is for displacements -MR/2 up to MR/2 and for thickness measure-
ment on 0 up to 2 MR (MR=Measuring range).
In case of minimum and maximum measured value is ‚0‘, the default scaling is used.
The minimum and maximum measured value is to output in millimeters. The available
output range of the analog output is then spread between the minimum and maximum
measured value. The minimum and maximum measured value must be between -120.0
and 120.0.
The minimum and maximum measured value is processed with four decimal places.