Page 108
Appendix |ASCII Communication with Controller
confocalDT 24x1
A 5.5.2.6 Set Video Output
OUTVIDEO NONE|[RAW] [DARK] [LIGHT] [DARKTAB] [LIGHTTAB] [THRES]
Setting the data to be transmitted in a video transmission.
-
NONE: No video signals
-
RAW: Output of the unconditioned signal
-
DARK: Output of the dark-corrected signal
-
LIGHT: Output of the light source corrected signal
-
DARKTAB: Output of the dark-correction table
-
LIGHTTAB: Output of the light-correction table
-
THRES: Output of the threshold table
Video signals can only be transmitted via the Ethernet interface.
A 5.5.3 Switching Outputs
A 5.5.3.1 Error Switching Outputs
ERROROUT1 NONE|ER1|ER2|ER12|LI1|LI2|LI12
ERROROUT2 NONE|ER1|ER2|ER12|LI1|LI2|LI12
Setting the error switching outputs.
-
NONE: No output on the error switching outputs
-
ER1: Switching output is switched in case of intensity error
-
ER2: Switching output is switched in case of a measured value outside of the measu-
ring range
-
ER12: Switching output is switched in case of an intensity error or a measured value
outside of the measuring range
-
LI1: Switching output is switched in case of deceeding the lower limit
-
LI2: Switching output is switched in case of exceeding the upper limit
-
LI12: Switching output is switched in case of deceeding the lower limit or exceeding
the upper limit
A 5.5.3.2 Limit Values
Controller IFC24x1
Controller IFC24x1MP
ERRORLIMIT DIST1|DIST2
|THICK12[<Unterer Grenzwert>
[<Oberer Grenzwert>]]
ERRORLIMIT DIST1|DIST2|
DIST3|DIST4|DIST5|DIST6|
THICK12|THICK13|THICK14|
THICK15|THICK16|THICK23|
THICK24|THICK25|THICK26|
THICK34|THICK35|HICK36|
THICK45|THICK46|THICK56
[<Unterer Grenzwert> [<Oberer
Grenzwert>]]
Selection of the signal and setting the limits for a limit consideration via the error switch-
ing outputs.
-
DIST1: The limits refer to the displacement 1
-
DIST2: The limits refer to the displacement 2 (only for thickness measurement)
-
THICK12: The limits refer to the difference between displacement 1 and 2 (only for
thickness measurement)
The limits are specified in millimeters with six decimals and must be between -120.0 and
120.0.