MI 3101 EurotestAT: Appendix A
Fuse table
100
A Appendix A - Fuse table
Fuse type NV
Disconnection time [s]
35m 0.1 0.2 0.4 5
Rated
current
(A)
Min. prospective short- circuit current (A)
2
32.5
22.3
18.7
15.9
9.1
4
65.6
46.4
38.8
31.9
18.7
6
102.8
70
56.5
46.4
26.7
10
165.8 115.3 96.5 80.7 46.4
16
206.9 150.8 126.1 107.4 66.3
20
276.8 204.2 170.8 145.5 86.7
25
361.3 257.5 215.4 180.2 109.3
35
618.1 453.2 374 308.7 169.5
50
919.2 640 545
464.2
266.9
63
1217.2 821.7 663.3 545 319.1
80
1567.2 1133.1 964.9 836.5 447.9
100
2075.3
1429
1195.4
1018
585.4
125
2826.3
2006
1708.3
1454.8
765.1
160
3538.2 2485.1 2042.1 1678.1 947.9
200
4555.5 3488.5 2970.8 2529.9 1354.5
250
6032.4 4399.6 3615.3 2918.2 1590.6
315
7766.8 6066.6 4985.1 4096.4 2272.9
400 10577.7 7929.1 6632.9 5450.5 2766.1
500 13619 10933.5 8825.4 7515.7 3952.7
630
19619.3 14037.4 11534.9 9310.9 4985.1
710
19712.3 17766.9 14341.3 11996.9 6423.2
800
25260.3 20059.8 16192.1 13545.1 7252.1
1000
34402.1 23555.5 19356.3 16192.1 9146.2
1250
45555.1 36152.6 29182.1 24411.6 13070.1
Fuse type gG
Disconnection time [s]
35m 0.1 0.2 0.4 5
Rated
current
(A)
Min. prospective short- circuit current (A)
2
32.5 22.3 18.7 15.9 9.1
4
65.6 46.4 38.8 31.9 18.7
6
102.8
70
56.5
46.4
26.7
10
165.8 115.3 96.5 80.7 46.4
13
193.1 144.8 117.9 100
56.2
16
206.9 150.8 126.1 107.4 66.3
20
276.8 204.2 170.8 145.5 86.7
25
361.3 257.5 215.4 180.2 109.3
32
539.1 361.5 307.9 271.7 159.1
35
618.1 453.2 374 308.7 169.5
40
694.2 464.2 381.4 319.1 190.1
50
919.2
640
545
464.2
266.9
63
1217.2 821.7 663.3 545 319.1
80
1567.2 1133.1 964.9 836.5 447.9
100
2075.3
1429
1195.4
1018
585.4