MI 3101 EurotestAT: Instrument operation
Automatic testing
34
Example:
Item/keys
Chapter
reference
Comment
Auto sequence, TEST
4.1
Selection of auto sequence operation in main
menu.
/
4.3.1
Selection of test sequence number
10.
F1
4.3.4
Enter into sequence name editing menu.
F1
4.3.4
Enter the sequence name editor.
Sock. 6A / 30mA(AC)
4.3.4
Enter the name of auto sequence.
F2
4.3.4
Accept name and exit into sequence name editing
menu.
4.3.4
Select description of test field.
F1
4.3.4
Enter the description of test editor.
Verification of wall
socket, protected with
fuse and RCD
4.3.4
Enter the description.
F2
Accept description and exit into sequence name
editing menu.
TEST
4.3.4
Exit sequence name editing menu.
4.3
Enter into sequence field.
/
4.3.2
Select
CONTINUITY.
TAB
4.3.2
Enter test parameter selection mode.
TEST
R200mA
Limit
ON
Limit
0.1
5.2
Set test parameters for equipotential bonding
resistance.
TAB
4.3.2
Exit parameter mode.
F1
4.3.2 Set
PAUSE
(wait to prepare for measurement).
/
4.3.7 Set
PAUSE:
ON.
4.3.7 Select
COMMENT.
4.3.7 Select
COMMENT:
#1.
F1
4.3.7 Enter
Edit comment
menu.
Disconnect mains,
univ. cable + ext.
4.3.7
Enter the comment.
F2
4.3.7
Save the comment.
TEST
4.3.7
Store the comment to location #1.
4.3.7 Select
COMMENT:
#2.
F1
4.3.7 Enter
Edit comment
menu.
Commander
4.3.7
Enter the comment.
F2
4.3.7
Save the comment.
TEST
4.3.7
Store the comment to location #2.
4.3.7 Select
COMMENT:
#3.
F1
4.3.7 Enter
Edit comment
menu.
Connect mains
4.3.7
Enter the comment.
F2
4.3.7
Save the comment.
TEST
4.3.7
Store the comment to location #3.
4.3.7 Select
COMMENT:
#4.
F1
4.3.7 Enter
Edit comment
menu.
Turn ON RCD
4.3.7
Enter the comment.
F2
4.3.7
Save the comment.
TEST
4.3.7
Store the comment to location #4.