MI 3101 EurotestAT: Instrument description
Display organization
18
Warning!
Phase voltage on the PE terminal! Stop the activity immediately
and eliminate the fault / connection problem before proceeding with any
activity!
3.5.2 Menu line
In the menu line the name of the selected function is displayed. Additional informations
about active cursor / TEST keys and battery condition are shown.
Function name.
Time.
Active keys on cursor /
TEST
keypad (
and
TEST
in this
example).
Battery capacity indication.
Low battery.
Battery is too weak to guarantee correct result. Replace or
recharge the battery cells.
Recharging in progress (if power supply adapter is connected).
3.5.3 Message field
In the message field different warnings and messages are displayed.
Warning!
High voltage is applied to the test terminals.
Measurement is running, consider displayed warnings.
Conditions on the input terminals allow starting the measurement (
TEST
key), consider other displayed warnings and messages.
Conditions on the input terminals do not allow starting the measurement
(
TEST
key), consider displayed warnings and messages.
Test leads resistance in
CONTINUITY
tests is not compensated, see
Chapter
5.2.3
for compensation procedure.
Test leads resistance in
CONTINUITY
tests is compensated.
Possibility to perform reference measurement (Zref) in
Δ
U
sub-function.
RCD tripped-out during the measurement (in RCD functions).
Instrument is overheated, the temperature inside the instrument is higher
than the safety limit, and measurement is prohibited until the temperature
decreases under the allowed limit.
Fuse F1 has blown or not inserted (
CONTINUITY
and
EARTH
functions).
It is possible to store result(s).
High electrical noise during measurement. Results may be impaired.