MI 3101 EurotestAT:
Technical specifications
97
Nominal voltage range ............................100 V
440 V
Nominal frequency ..................................50 Hz
Maximum test current (at 400V)..............267 A (10 ms)
Calculation of prospective short-circuit current (standard voltage value):
Z
V
230
=
I
K
U
L-N
= 230 V
10 %
Z
V
400
=
I
K
U
L-L
= 400 V
10 %
Calculation of prospective short-circuit current (non-standard voltage value):
L
-
L
L)
-
L
(
N
MAX
ph
3
KMAX
Z
2
×
3
U
×
C
=
I
HOT
L)
-
(L
L)
-
L
(
N
MIN
ph
3
KMIN
Z
2
×
3
U
×
C
=
I
L
-
L
L)
-
L
(
N
MAX
ph
2
KMAX
Z
U
×
C
=
I
HOT
L)
-
(L
L)
-
L
(
N
MIN
ph
2
KMIN
Z
U
×
C
=
I
N
-
L
N)
-
L
(
N
MAX
N)
-
(L
KMAX
Z
U
×
C
=
I
HOT
N)
-
(L
N)
-
L
(
N
MIN
N)
-
(L
KMIN
Z
U
×
C
=
I
2
L
-
L
2
L
-
L
L
-
L
X
+
R
=
Z
(
)
2
L
-
L
2
L
-
L
HOT
L)
-
(L
X
+
R
×
5
.
1
=
Z
2
N
-
L
2
N
-
L
N
-
L
X
+
R
=
Z
(
)
2
N
-
L
2
N
-
L
HOT
N)
-
(L
X
+
R
×
5
.
1
=
Z
U
N(L-N)
= 230 V
10 %
U
N(L-L)
= 400 V
10 %
230 V < U
N
< 400 V
C
MAX
1.05
1.10
C
MIN
0.95
1.00
8.9.2 High precision fault loop impedance
Measuring range according to EN61557 is 5.0
1999 m
Measuring range (m
) Resolution
(m
)
Accuracy
0.0
199.9
0.1
200
1999
1
(5 % + 1 m
)
Nominal voltage range ............................100 V
440 V
Nominal frequency ..................................50 Hz
Maximum test current (at 230 V).............154 A (10 ms)
Calculation of prospective short-circuit current (standard voltage value):
Z
V
230
=
I
K
U
L-PE
= 230 V
10 %
Calculation of prospective short-circuit current (non-standard voltage value):
PE
-
L
PE)
-
L
(
N
MAX
PE)
-
(L
KMAX
Z
U
×
C
=
I
HOT
PE)
-
(L
PE)
-
L
(
N
MIN
PE)
-
(L
KMIN
Z
U
×
C
=
I
2
PE
-
L
2
PE
-
L
PE
-
L
X
+
R
=
Z
(
)
2
PE
-
L
2
PE
-
L
HOT
PE)
-
(L
X
+
R
×
5
.
1
=
Z