MI 3101 EurotestAT: Maintenance
Replacing fuses/Cleaning/Calibration/Service
89
7 Maintenance
Unauthorized person is not allowed to open the EurotestAT instrument. There are no user
replaceable components inside the instrument, except three fuses and batteries under rear
cover.
7.1 Replacing fuses
There are three fuses under back cover of the EurotestAT instrument.
F1
M 0.315 A / 250 V, 20
5 mm
This fuse protects internal circuitry of continuity function if test probes are
connected to the mains supply voltage by mistake during measurement.
F2, F3
F 4 A / 500 V, 32
6.3 mm
General input protection fuses of test terminals L/L1 and N/L2.
Warnings:
Disconnect any measuring accessory and power off the instrument before
opening battery/fuse compartment cover, hazardous voltage inside!
Replace blown fuse with original type only, otherwise the instrument may be
damaged and/or operator’s safety impaired!
Position of fuses can be seen in
Figure 3.4
in chapter
3.3 Back panel
.
7.2 Cleaning
No special maintenance is required for the housing. To clean the surface of the instrument
use a soft cloth slightly moistened with soapy water or alcohol. Then leave the instrument
to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!
7.3 Periodic calibration
It is essential that the test instrument is regularly calibrated in order technical specification
listed in this manual is guaranteed. We recommend an annual calibration. Only an
authorized technical person can do the calibration. Please contact your dealer for further
information.
7.4 Service
For repairs under warranty, or at any other time, please contact your distributor.