MI 3101 EurotestAT: Measurements
Continuity
50
5.2.3 Compensation of test leads resistance
This chapter describes common principle for compensation of test leads resistance for
both CONTINUITY functions. The compensation is required to eliminate the influence of
test leads resistance plus internal resistances of the instrument. The lead compensation is
very important to obtain correct result. The compensation status (
/
) is indicated
in the message field.
Key:
F1
Enters test leads resistance compensation menu for any of mentioned
functions.
See chapter
4.2 Single test
for functionality of keys.
Figure 5.10: Test leads resistance
compensation menu
Keys:
TEST
Performs compensation
/
Sets function to be compensated
The instrument compensates following Continuity subfunctions.
Compensation NPE
Same compensation for both
7 mA
and
200
mA
measurements.
Short N and PE terminals.
Compensation LPE
Same compensation for both
7 mA
and
200
mA
measurements.
Short L and PE terminals