MI 3101 EurotestAT: Appendix G
ES1 – application of regulative UNE-202008
128
See chapter
4.2 Single test
for functionality of keys.
Figure G.5: Continuity
Test parameters for resistance measurement
TEST Resistance
measurement
sub-function
[R200mA NPE, R7mA NPE, R200mA
LPE, R7mA LPE, R-PE(loop)]
If
R-PE(loop) then
Fuse I
Rated current
of fuse [---, 6 A, 10 A, 16 A, 20 A, 25 A]*.
Uc Touch
voltage
limit
[25 V, 50 V].
RPE_lim
Maximum
PE resistance.
RPE_cal
Calibrated value
of PE resistance given by reference loop impedance
measurement (see
G.3.3 Fault loop impedance and prospective fault current
).
else
Limit
Maximum resistance
[OFF, 0.1
÷ 20.0
]
*--- Means no fuse selected.
G.3.2.1 Continuity R200 mA measurement
See
Chapter 5.2.1 Continuity R200 mA measurement
.
G.3.2.2 7 mA resistance measurement
See
Chapter 5.2.2 7 mA resistance measurement
.