MI 3101 EurotestAT: Instrument operation
Miscellaneous
43
4.4.6 Communication port
The communication port (RS232 or USB) can
be selected in this menu.
Figure 4.31: Communication port selection
Keys:
/
Select
communication
port.
TEST
Confirms selected port.
ESC
Exits without changes.
Note
:
Only one port can be active at the same time.
4.4.7 Locator
This function enables tracing electrical lines.
Keys:
TEST
Starts locator function.
ESC
Exits miscellaneous menu.
See chapter
5.10 Locator
for locator operation.
4.4.8 Operator
This menu enables registering the operator of
the instrument. Selected operator name
appears on the bottom of the LCD during
turning on of the instrument. It is also
associated to stored measurement results.
Up to 5 operators can be defined.
Figure 4.32: Operator menu
Keys:
/
Select operator.
TEST
Accepts selected operator.
ESC
Exits to miscellaneous menu without changes.
F1
Enters operators name edit menu.