USER MANUAL, RSS-1000-CVSI
7000-5970
Rev.
A
page
378
of
396
18. Enter 0.2mm in the Z-Axis fine adjustment and carefully position the end effector to the calibration
fixture height, adjusting the values as needed until the cartridge gently contacts the fixture.
18.1. Please note that the orientation of the cartridge will likely no longer be aligned at this point
and/or out of position in the X-Axis. DO NOT loosen set screw of the Cartridge Lock to
account for orientation or position issues (see next step.)
19. Slide the Spring Retainer of the CV-H8-AV down until a sufficient amount of pressure has been
applied to the End Effector Cartridge Holder.