14. Insert CV Cartridge and Cartridge Lock into the End Effector Cartridge Holder. Rotate clockwise
to secure in place.
15. Tap the Home Button
to ensure the End Effector is positioned at 0,0,0,0 (X, Y, Z, Θ)
16. Slowly and carefully jog the End Effector down in the Z-Direction
towards the calibration
fixture, stopping at roughly 5mm above the fixture.
17. Tap the Motion Control button
to toggle to the fine adjustment menu.