Please enter a valid month.
Displays when an operator enters an invalid
month. Enter a valid month.
Change flow mode?
Select
Yes
to change the flow mode to mL/m.
ISI interface module failure. Disable ISI to
proceed.
The ISI interface is not functioning properly.
Disable ISI at the injector system to continue
with the injection.
ISI not synchronized
The ISI remote start is already on when the
injector is armed, or the ISI hand switch disable
is turned on after an injection is started using
the injector hand switch or foot switch.
ISI not ready
ISI is not communicating correctly with the
system. Check ISI connections and
communication.
Table 13 - 2: Popup Messages
Popup Message
Description/Resolution
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......