Specifications
17 - 131
17.3 ISI Technical Specifications
This section outlines the
Mark 7 Arterion
specifications for the output signals and input signals for the
ISI, and shows pinouts for legacy ISI connector, universal cable, and Siemens cable.
NOTE:
Systems are configured for Universal ISI operation. If configuring for Siemens sys-
tem, contact local service for assistance.
17.3.1 ISI Output Specifications
The tables below list the output signals and the relay contact outputs for the
Mark 7 Arterion
ISI.
Table 17 - 1: Output Signals
Signal Name
Description
Extended_Arm
This signal is a pair of relay contacts output from the Injector. When active,
it indicates that the Injector is in an armed state and is ready to accept the
start input from the Imaging System. This signal operates as an Extended
Arm; it becomes active with the arming of the Injector and remains active
after an injection until the signal that initiated the injection (e.g.
INJ_START signal from the Imaging System) is deactivated.
NOTE:
The Injector still disarms at the completion of the
injection; only this ISI armed signal is extended.
Injecting
This signal is a pair of relay contacts output from the Injector. It indicates
that the piston on the Injector is in motion.
X-Ray Trigger
This signal is a pair of relay contacts output from the Injector. When the
Injector X-ray delay has expired, this signal becomes active and notifies
the Imaging System to start the X-ray. Like the Extended Arm signal, the
X-RAY_TRIGGER signal remains active until the signal initiating the
injection (e.g. INJ_START signal from the Imaging System) is deactivated.
Inj_Hand_Switch_On
This signal is a pair of relay contacts output from the Injector. When the
INJ_HAND_SWITCH_DISABLE input is active and the Injector is armed,
this signal indicates when the Injector hand switch (contrast activation
only) or foot switch is in an “on” position; otherwise, this signal is inactive.
Table 17 - 2: Relay Contact Outputs
2A @ 28VDC or 0.5A @ 125VAC
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......