MEDRAD
®
Mark 7 Arterion Injection System
9 - 42
NOTE:
The displayed values are based on the last used protocol or the default values.
2.
Select the box corresponding to a parameter to change it.
3.
Use the numeric keypad to enter the protocol parameter.
4.
Commit the value by selecting
Enter
or another parameter.
5.
Select
Delay
. A numeric keypad displays with
X-Ray
and
Inject
buttons.
a.
Enter the delay time. Setting delay time to zero is equivalent to having no delay.
b.
Select
X-Ray
or
Inject
.
c.
Select
Enter
to commit the value.
NOTE:
A programmed delay parameter only functions when using ISI.
6.
Repeat steps
to change additional parameters.
9.1.2 Set Injection Parameters on Home Tab - Phased
Phased Protocols have up to four different sets of flow rates and volumes for single continuous
injection. Operators can enter up to four phases per protocol. The pressure limit remains consistent for
each phase.
NOTE:
The initial rise time is dependent on the value the operator enters. The intra-phase
rise time is fixed.
NOTE:
If the Phased option is not visible, go to the Display Control Unit
Options
tab to
enable Phased injections. See
NOTE:
ISI does not function with Phased protocols.
1.
Select the
Phased
tab (A).
Figure 9 - 2: Set Phased Protocol
From the
Programmed
window, operators can modify protocol parameters and existing
phases, add phases, or delete phases.
•
To modify an existing phase, go to step
•
To modify the pressure limit, rise time, or delay, go to step
•
•
.
2.
To modify values for an existing Phase, select the
Flow Rate
or
Volume
to the right of the
index number (B) for that phase.
a.
Select either
Flow Rate
or
Volume
.
b.
Use the numeric keypad to enter the protocol parameter.
c.
Commit the value by selecting
Enter
or another parameter.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......