MEDRAD
®
Mark 7 Arterion Injection System
7 - 34
7.2 Injector Head Position
The
Mark 7 Arterion
Injector Head contains a sensor that monitors the head’s position: Purge (upright)
(X), Intermediate (Y), or Inject (downward) (Z). The head position determines how the data displays and
the available functions. Use the handle and the back of the Injector Head (but not the Manual Knob) to
rotate the head into position.
Figure 7 - 2: Injector Head Position
7.3 Syringe Interface
The syringe interface supports a single 150 mL syringe. Operators can attach a syringe to the head
from the front of the injector (front loading). The syringe locks into place when the drop front is fully
Table 7 - 1: Injector Head Position and Functionality
Functions
Injector Head Position
Purge (X)
Intermediate (Y)
Inject (Z)
Fill
Enabled
Enabled
Enabled
Inject
Disabled
Disabled
Enabled
Z
X
Y
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......