16 - 102
Figure 16 - 20: Tightening the Clamp Knob
4.
Using the 3/16” Hex Key provided, turn the adjusting screw (A) clockwise until Mounting
Lever (B) contacts the bottom of the rail.
Figure 16 - 21: Adjust Screw and Mounting Lever
5.
Turn the same adjusting screw (A) counterclockwise approximately 45 degrees; remove the
Hex Key.
Figure 16 - 22: Table Mount Knob
6.
Tighten the Clamp Knob until the Bracket Mount is firmly attached to the rail.
7.
Loosen the knob (C) by turning as far as possible counterclockwise.
8.
With the Display Control Unit attached to the Fulcrum Mount, insert the Fulcrum Mount pin
into the top of the Fixed Table Mount bracket.
9.
Secure the Display Control Unit by turning the knob (C) clockwise. When tightened properly,
the Display Control Unit should not move when pressing the touch screen.
A
B
C
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......