System Overview
This chapter describes:
•
•
•
"System Technical Specifications"
•
"High Pressure Connector Tubing Specifications (Non-Twist & Go)"
•
•
•
•
•
•
Figure 5 - 1: Mark 7 Arterion Injection System
5.1 Injection Protection
The following means are provided to protect against over and under injections:
An on-screen indication of insufficient volume is provided whenever the total volume programmed to
be delivered is greater than the amount of fluid in the syringe.
The system monitors injections to detect over rate or over volume conditions due to system faults. The
delivered volume is also monitored against the total programmed volume for the injection.
Once the system has disarmed a tone will sound and a disarm message displays on the Display Control
Unit screen.
When any fault condition is detected, the injection will stop.
A
Display Control Unit
B
Injector Head
C
Power Unit
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......