6 - 32
6.6 Display Control Unit Lock-outs
The Display Control Unit will be locked-out while an operator is interacting with the Injector Head
controls, or while an operator is accessing another Display Control Unit in a dual Display Control Unit
system.
The inactive Display Control Unit remains in lock-out until (A):
1.
the operator finishes accessing the active Display Control Unit, and active Display Control
Unit displays the Home tab,
2.
or, an operator has accessed the active Display Control Unit and doesn’t do anything for a
minute,
3.
or, an operator has not interacted with the Injector Head controls for a period of several
seconds.
Figure 6 - 2: Display Control Unit Locked Out
6.7 Performing Touch Screen Calibration
Calibrate the Display Control Unit if the touch screen does not respond appropriately when pressing
buttons on the screen. To calibrate at the Safety Screen, simultaneously press the
Brightness Up
and
Brightness Down
buttons on the back of the Display Control Unit. Follow the on-screen instructions
that display.
NOTE:
Touch the center of the calibration targets to ensure proper touch screen
calibration.
Figure 6 - 3: Touch Screen Calibration
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......