Arming and Injecting
11 - 63
Single mL/m and Phased protocols:
select
Arm
(B).
Figure 11 - 2: Phased protocol
3.
Visually confirm that all air has been purged from the syringe and disposable set, and select
Yes
.
NOTE:
There is sufficient volume remaining in the syringe. For Single mL/s and mL/m
protocols with insufficient volume remaining in the syringe for the programmed
protocol, the system gives the option of overriding the programmed volume and
using the available volume remaining. If the operator selects
Yes
, the system arms
with the new programmed volume. If the operator selects
No
, the system will not
arm. Adjust the protocol volume to be equal or less than the volume remaining in
the syringe. For Phased protocols with insufficient volume remaining in the syringe
for the programmed protocol, the system will not arm.
NOTE:
If the injector is set in the mL/m mode, a confirmation popup displays confirming
that the injection will be performed using mL/m.
NOTE:
The system stays armed until:
• the operator presses
Disarm
.
• any Injector Head button is pressed.
• a reverse piston motion of greater than 2 mL occurs via the Manual Knob.
• the drop front is lowered.
• the Injector Head is rotated out of the Inject position.
• a start switch is connected or disconnected.
• a 30 minute time-out occurs.
• ISI signals a disarm.
4.
The Armed Light illuminates solid and the system is ready for injection. For more information,
refer to
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......