Arming and Injecting
11 - 65
NOTE:
The system will stay armed until:
• the user presses
Disarm
.
• any Injector Head button is pressed.
• a reverse piston motion of greater than 2 mL occurs via the manual knob.
• the drop front is lowered.
• the head is rotated out of the Inject position.
• a start switch is connected or disconnected.
• a 30 minute time-out occurs.
• ISI signals a disarm.
4.
The Armed Light illuminates solid and the system is ready for injection. For more information,
refer to
"11.3-Performing an Injection"
11.3 Performing an Injection
The
Mark 7 Arterion Injection System
can perform Fixed Flow Rate and Variable Flow Rate injections.
For Fixed Flow Rate, Arm Single mode injections, the injector disarms after the injection is complete or
an operator releases the hand switch, foot switch, or Imaging System start switch. For Fixed Flow Rate,
Arm Multi mode injections and Variable Flow Rate injections, the injector remains armed until one of
the disarm criteria are met.
NOTE:
Ensure the fluid path is open before performing an injection.
11.3.1 Performing a Single mL/s Injection in Arm Single Mode
NOTE:
The term start switch in this section is used to refer to the hand switch, foot switch
or hand controller.
1.
Press and hold the start switch to start the injection, and continue holding until the injection
is complete.
The system stops injecting and disarms when:
•
the programmed volume is delivered, or
•
the operator releases the start switch, or
•
the operator presses the Display Control Unit screen or Injector Head controls.
During the injection, the Injection Indicator displays below the syringe graphic and the Armed
Light flashes.
2.
"11.4-Completing an Injection"
.
11.3.2 Performing a Single mL/m Injection in Arm Single Mode
NOTE:
The term start switch in this section is used to refer to the hand switch, foot switch
or hand controller.
1.
Press and release the start switch to start the injection.
The system stops injecting and disarms when:
•
the programmed volume is delivered, or
•
the operator presses and releases the start switch again, or
•
the operator presses the Display Control Unit screen or Injector Head controls.
During the injection, the Injection Indicator displays below the syringe graphic and the Armed
Light flashes.
2.
"11.4-Completing an Injection"
.
11.3.3 Performing a Single mL/s or Variable Flow Rate Injection in Arm Multi Mode
NOTE:
The term start switch in this section is used to refer to the hand switch, foot switch
or hand controller.
1.
Press and hold the start switch to start the injection, and continue holding until the injection
is complete.
The system stops injecting and remains armed when:
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......