MEDRAD
®
Mark 7 Arterion Injection System
10 - 56
Figure 10 - 3: Twist & Go Connection
3.
Turn the (A) Luer nut clockwise to securely tighten the high pressure connector tubing to the
syringe tip (B).
4.
Purge all air from the high pressure connector tubing. Turn the Manual Knob clockwise to
push contrast out until all of the air bubbles have been removed from the High Pressure
Connector Tube. Gentle tapping at any connection point may be needed to facilitate air
removal.
10.5 Installing the MEDRAD
®
VFlow
Hand Controller
1.
Ensure that the injector is in the disarmed state.
2.
Open the hand controller bag using sterile technique.
3.
Using sterile technique, remove the hand controller from the bag.
4.
Plug the hand controller into the underside of the Injector Head (see drawing below). Listen
for an audible “click” to confirm proper connection.
Figure 10 - 4: Hand Controller Connector
NOTE:
If the hand controller is damaged or does not work properly, discontinue use and
discard.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......