13 - 75
13
System Messages
The
Mark 7 Arterion Injection System
displays Sentinel Messages and Popups to alert the operator that
action is required.
This chapter describes:
•
•
•
13.1 Error Messages
Error messages are system malfunction messages which require power to be removed from the
system. Error messages are accompanied by three beeps. Some error messages provide suggestions
to prevent the condition from recurring. If the condition cannot be corrected, record the code and
number from the lower left corner of the dialog box, then call Bayer HealthCare Services for assistance.
Error messages are divided into categories of function level. Each category is also divided into specific
errors. Below is a list of the categories and suggested repair sequences. These are to be tried in order,
not to be performed all at once. For further assistance, contact Bayer HealthCare Services or an
authorized dealer.
NOTE:
Before replacing any parts, cycle power to the system. This initiates a system self-
test. If this does not correct the problem, replacement of serviceable components
as needed may be required.
13.2 Sentinel Messages
Sentinel messages display in the Sentinel window located in the lower right of the touch screen. A
Sentinel message informs an operator of items that need attention. If multiple items need to be
corrected, the Sentinel window shows a list of the items that need to be corrected. As each item is
corrected, the system removes the corresponding message from the Sentinel window.
Electric Shock Hazard - Serious patient and/ or worker injury or death may result
.
•
Perform calibration when specific components listed in “Disassembly/Assembly &
Replacement Parts” are replaced in the system.
Mechanical Hazard - Minor or moderate patient and/or worker injury may result.
•
Remove power and disconnect patient when system malfunction occurs.
CAUTION
CAUTION
Table 13 - 1: Sentinel Messages
Sentinel Message
Description/Resolution
Attach hand controller
Attach hand controller
Attach start switch or enable ISI
Attach a start switch or enable ISI.
Install Syringe
Install a syringe.
Close Drop Front
Completely close the drop front.
Advance Plunger
Fully advance the plunger in the syringe.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......