MEDRAD
®
Mark 7 Arterion Injection System
9 - 46
Figure 9 - 6: Index Number
c.
To enter a new phase, select an empty index number (C). A new phase is added with
default values for flow rate and volume.
d.
Enter the phase values, as needed
NOTE:
The currently selected values update in the top row parameters. The value in the
Phased Selection pane does not update until the value is committed.
e.
To commit the values, select a different parameter.
f.
Repeat this step for each phase to be added.
6.
Select
Pressure
or
Rise Time
.
a.
Enter the parameter values.
b.
To commit the values, select a different parameter.
c.
Repeat this step for each parameter to be set.
NOTE:
The operator enters the initial rise time. The intra-phase rise time is fixed.
7.
Select
Save
.
9.2.1.3 Create a New Variable Flow Rate Protocol on the Protocols tab
A Variable Flow Rate protocol injects a set Volume of contrast at a Flow Rate determined by the Hand
Controller. As an operator depresses the Hand Controller plunger, the system increases the flow rate
until it reaches the maximum Flow Rate set by the operator.
NOTE:
If the Variable Flow Rate option is not visible, go to the Display Control Unit
Options
tab to enable MEDRAD VFlow. See
NOTE:
ISI does not function with Variable Flow Rate protocols.
1.
Select the
Protocols
tab.
2.
Select the
Type
button until Variable is highlighted. A list of Variable Flow Rate protocols
displays.
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......