M133C Electric Power/Energy Calibrator
MEATEST
101
Operation manual
If query is sent, calibrator returns the value of modulation frequency using standard exponential
format. Example: 5Hz is returned as 5.000.
Example:
Modulation frequency 5.3Hz:
PHAR:FREQ:MOD 5.3
Notes:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PIH
This command switches the calibrator in Power Interharmonic mode.
Example:
PIH
[SOUR]:PIH:VOLT<n>(?) <DNPD>
This command sets the output voltage value in channel <n>
<DNPD>
Parameter represents the value of the output voltage expressed in V.
<n>
Parameter represents the output channel (1, 2 or 3).
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V in channel 1:
PIH:VOLT1 230.5
Note:
Command switches the calibrator in Power Interharmonic mode.
[SOUR]:PIH:VOLT<n>:PHAS(?) <DNPD>
This command sets the phase angle between voltage output in channel <n> and internal reference
phase.
<DNPD>
Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential
format. Example: 156.3
is returned as 1.002.
Example:
PIH:VOLT1:PHAS 0