M133C Electric Power/Energy Calibrator
MEATEST
99
Operation manual
Example:
Set harmonic order 2 to 5.3% in voltage channel 1:
PHAR:CURR1:HARM2 5.3
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:HARM<y>:PHAS(?) <DNPD>
This command sets the phase shift of <y> harmonic component in channel <n>
<DNPD>
Parameter represents the phase shift of <y> harmonic component expressed in °.
<n>
Parameter represents the output channel (1, 2 or 3).
<y>
Parameter represents the harmonic order (2 to 50). Order 1 has zero phase shift and can’t be
changed.
If query is sent, calibrator returns the phase shift of harmonic component using standard
exponential format. Example: 10° is returned as 1.001.
Example:
Set phase shift of harmonic component 2 to 5.3° in current channel 1:
PHAR:CURR1:HARM2:PHAS 5.3
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:MOD(?) <DNPD>
This command sets the modulation level for current output in channel <n>.
<DNPD>
Parameter represents the modulation level in %.
If query is sent, calibrator returns the modulation level using standard exponential format.
Example: 12.45
is returned as 1.001.
Example:
PHAR:CURR1:MOD 12.45
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:MOD:SHAP(?) <CPD> { SIN | RECT | OFF }
This command sets type of modulation for current output in channel <n>.