MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
108
[SOUR]:PDIP:TIME(?) <DNPD>,<DNPD>,<DNPD>,<DNPD>,<DNPD>
This command sets the time periods t1, t2, t3, t4 and t5 in Power Dip/Swell mode.
<DNPD>,<DNPD>,<DNPD>,<DNPD>,<DNPD>
Parameters represent the value of times (t1, t2, t3, t4 and t5) expressed in s.
If query is sent, calibrator returns time values using standard exponential format. Example: 50s is
returned as 5.001.
Example:
period t1 100ms
period t2 20ms
period t3 30s
period t4 200ms
period t5 0s
PDIP:TIME 0.1, 0.02, 30, 0.2, 0
Notes:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:TRIG:INP(?) <CPD> { OFF | IN3 }
This command enables (disables) external trigger for Dip/Swell mode.
OFF – disables external trigger for Dip/Swell function
IN3 – enables external trigger for Dip/Swell function (TTL falling edge applied to input
IN3)
If query is sent, calibrator returns IN3 if external trigger is enabled or OFF if it is disabled.
Example:
PDIP:TRIG:INP IN3
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:TRIG:SYNC(?) <CPD> { OFF | ON }
This command enables (disables) synchronization of Dip/Swell shape.
OFF – Dip/Swell shape starts immediately after trigger
ON – start of Dip/Swell shape is synchronized with the phase 0°
If query is sent, calibrator returns ON if synchronization is enabled or OFF if it is disabled.
Example:
PDIP:TRIG:SYNC ON
Note:
Command switches the calibrator in Power Dip/Swell mode.