M133C Electric Power/Energy Calibrator
Operation manual
78
MEASure a CONFigure subsystem
This subsystem allows controling the internal multimeter.
Keyword
Parameters
MEASure?
CONFigure(?)
<CPD> { VOLTage | CURRent | FREQuency }
MEAS?
This query returns value measured by the multimeter.
calibrator returns the set value of power using standard exponential format. Example: 5.456 V is
returned as 5.000.
Example:
MEAS? Calibrator returns measured value
CONF(?) <CPD> { VOLT | CURR | FREQ }
This command sets the function of internal multimeter.
If query is sent, calibrator returns VOLT, CURR or FREQ according to selected multimeter
function.
Example:
CONF VOLT sets the multimeter to function voltage DC
SOURce subsystem
This subsystem allows controlling the individual functions of the calibrator.
Keyword
Parameters
[SOURce]
:MODE? { PAC | PDC | PACI |
PDCI
| PACE |
PHAR | PIH | PDIP | PDCE
| EAC | EDC |
EACI | EDCI | VAC | VDC | CAC | CDC | CACI | CDCI }
:PAC
:POWer(?)
<DNPD>
[:POWer]
:UNIT(?)
<CPD> { W | VA | VAR }
:VOLTage(?)
<DNPD>
:CURRent(?)
<DNPD>
[:CURRent]
:PHASe(?)
<DNPD>
:POLarity(?)
<CPD> { LEAD | LAG }
:FREQuency(?)
<DNPD >
:PDC
:POWer(?)
<DNPD>