M133C Electric Power/Energy Calibrator
MEATEST
135
Operation manual
IEEE488.2 Common Commands
I/D (instrument identification)
*IDN?
This command returns the identification of the manufacturer, model, serial number and firmware
revision.
The reply is formatted as follows:
MEATEST,M133C,100002,1.22
Operation complete
*OPC
This command sets the OPC bit in the ESR (Event Status Register) when all pending operations
are complete.
Operation complete?
*OPC?
This command returns “1” to the output queue after all pending operations are complete.
Option identification?
*OPT?
This command return the instrument’s harware fitment.
The reply is formatted as follows:
1,a2,a3,0,0,0,0
Where:
1 .. indicates that the main unit (channel #1) is fitted –must be
a1 … indicates that additional unit (channel #2) is fitted
a2 … indicates that additional unit (channel #3) is fitted
0 … position reserved for future use.
a1 = 0 … unit channel #2 is not fitted
a1 = 1 … unit channel #2 is fitted
a2 = 0 … unit channel #3 is not fitted
a2 = 1 … unit channel #3 is fitted