MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
32
1.
Setting the amplitude (RMS value) of voltage (current)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
(current) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
2.
Setting the phase
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the phase
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ° softkey. The
value can be set using cursor buttons or rotary knob as well.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
3.
Setting the amplitude (RMS value) of Dip/Swell
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the Dip/Swell
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
4.
Setting Dip/Swell time periods
Repeatedly press the “Sel” button (in the middle of cursor buttons) until reuiered time
period (t1, t2, t3, t4 or t5) appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ms (s) softkey.
The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new time is set.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.