AC current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed (µA)
A-AC
300.0 mA
0.3
55
105 μ
A-AC
1.0 A
1.0
55
350 μ
A-AC
1.0 A
1.0
800
500 μ
A-AC
2.0 A
2.0
55
800 μ
A-AC
5.0 A
5.0
55
1.75 m
A-AC
10.0 A
10.0
55
4.5 m
A-AC
30.0 A
30.0
55
15.0 m
Table VII
AC/DC power test
Function
Range
Value (W)
Frequency (Hz) Deviation
allowed (W)
P 1
480W
480
55
0.216
P 0,5LA
480W
240
55
0.180
P 0,5LE
480W
240
55
0.180
P 1
2400W
2400
55
1.27
P 0,5LA
2400W
1200
55
0.972
P 0,5LE
2400W
1200
55
0.972
Table VIII
Voltage = 240V
Frequency test
Function
Range
Value (Hz)
Frequency (Hz) Deviation
allowed (Hz)
FREQ
10 kHz
10000.0
0.5
Table IX
Multimeter test
Function
Range
Value
Frequency (Hz) Deviation
allowed
FREQ
10 kHz
10000.0 Hz
0.5 Hz
V-DC
10 V
10.0 V
1.3 mV
V-DC
10 V
-10.0 V
1.3 mV
A-DC
20 mA
0.02 A
3.3 μA
A-DC
20 mA
-0.02 A
3.3 μA
Table X